Here is the procedure:

1. In Machine Setup add the rack or racks you are using.

rack1.jpg
rack1.jpg (70.98 KiB) Viewed 770 times

2. Calibrate the rack(s)

rack2.jpg
rack2.jpg (49.79 KiB) Viewed 770 times

3. Configure the Probe Station (rack) - For an FCR typr rack you can change the adapter types (empty for SM modules, PA25-SH for SP25 Stylus Holders, and PA25-20 for TP20's)

probe 4.jpg
probe 4.jpg (50.18 KiB) Viewed 770 times

Now go to the Probe Assy Dialog to correlate each Probe Assy with a Rack Port. First add the SP25M an be sure to click Update Assembly.

probe 5.jpg
probe 5.jpg (128.37 KiB) Viewed 770 times

4. Now add the SM25-1/SH25-1 or similar scanning module. You should see a dialog dislayed with an option to Assign to Port. Assign to Port 1

probe 3.jpg
probe 3.jpg (16.52 KiB) Viewed 770 times

5. Now at the bottom of the Probe Assy dialog assign the entire Probe to Port 2.

probe 6.jpg
probe 6.jpg (31.71 KiB) Viewed 770 times

Finally add another Probe Assembly and repeat steps 4 & 5. Be sure to assign the SM25-1 module to port 1 and the Probe Assy to Port 3.

This will allow changing just the Stylus Holder when changing from Probe Assy 0 to Probe Assy 1 while still using a common SM25 Scanning Module.

probe 7.jpg
probe 7.jpg (130.59 KiB) Viewed 770 times

 